Abstract
Test-retest reliability of lower extremity biomechanics by three-dimensional gait analysis
  
DOI:
EN KeyWords: three-dimensional gait analysis system  lower limbs  kinetics  kinematics  ground reaction force  test-retest reliability
Fund Project:国家自然科学基金青年项目(81503592,81503598);上海市卫生和计划委员会科研课题青年项目(20154Y0176)
作者单位
王铁强 上海中医药大学附属曙光医院骨伤科研究所上海 201203 
王晶 上海中医药大学附属曙光医院骨伤科研究所上海 201203 
张旻 上海中医药大学附属曙光医院骨伤科研究所上海 201203 
王学宗 上海中医药大学附属曙光医院骨伤科研究所上海 201203 
赵咏芳 上海中医药大学附属曙光医院骨伤科研究所上海 201203 
View Counts: 8727
PDF Download Counts: 5916
EN Abstract:
  Objective: To determine the test-retest reliability of kinetics, kinematics, ground reaction force (GRF) and the surface electromyography (sEMG) signal during gait in healthy subjects. Method: Kinetics, kinematics, GRF and the sEMG signals of 13 healthy subjects were collected by VICON (NEXUS 1.8.5) three dimensional gait analysis system, AMTI force plates and NORAXON wireless sEMG. The indicators of test retest reliability of kinetics, kinematics, GRF and the sEMG signals were intraclass correlation coefficient (ICC) and standard error of measurement (SEM), and they were used to analyze the test-retest reliability between those two sessions.Results: The test-retest reliability of walking speed, kinetics, kinematics, GRF and the sEMG signals was modest, and the value of ICC was from 0.78 to 0.96. The SEM % of kinematics, kinetics, GRF, and sEMG ranged from (4.18-15.6), t3.31-21.82, 1.70-16.67, and 8.00-11.11 respectively. Conclusion: The three-dimensional gait analysis and sEMG systems were presented to be a reliable and valid method for measuring kinetics, kinematics, GRF and the sEMG signal during gait.